The Stryker L10 LED Light Source with Advanced Imaging Modality (AIM) Technology is a sophisticated device designed to provide illumination during surgical procedures, offering modes such as White Light, ENV (Endoscopic Near-Infrared Visualization), and IRIS (Infrared Illumination System). While the device is engineered for reliability, users may occasionally encounter error codes, typically prefixed with an “E” (e.g., E-2), indicating specific issues that need attention. Below is an overview of common “E” errors and their potential solutions.
Common “E” Error Codes and Solutions
⚠️ E-2 Error: Light Source Not Emitting Light
- Possible Causes:
- Disconnected or improperly connected light cable.
- Faulty LED module or internal hardware malfunction.
- Overheating due to prolonged use or blocked ventilation.
- Solutions:
- Ensure the light cable is securely connected to both the light source and the endoscope.
- Inspect the light cable for any signs of damage; replace if necessary.
- Turn off the device and allow it to cool down if overheating is suspected.
- If the issue persists, contact Stryker technical support for further assistance.
⚠️ E-3 Error: ENV Mode Activation Failure
- Possible Causes:
- Incompatible or improperly connected AIM SafeLight™ cable.
- Incorrect or missing connection to the 1588 AIM Camera Control Unit (CCU).
- Software incompatibility or outdated firmware.
- Solutions:
- Verify that the AIM SafeLight™ cable is properly connected and compatible with the device.
- Ensure a USB A-to-A cable connects the L10 Light Source to the 1588 AIM CCU.
- Check for firmware updates and apply them if available.
- Consult the user manual for detailed setup instructions or contact Stryker support.
⚠️ E-4 Error: IRIS Mode Activation Failure
- Possible Causes:
- IRIS Ureteral Kit not connected or improperly connected.
- Use of non-Stryker fiber optic light guides.
- Device not in Standby mode during setup.
- Solutions:
- Connect only the Stryker IRIS Ureteral Kit to the IRIS connectors on the front panel.
- Ensure the device is in Standby mode before connecting or disconnecting the IRIS fibers.
- Avoid using third-party or non-approved accessories.
- Refer to the user manual for proper IRIS mode setup procedures.
⚠️ E-5 Error: Overheating Detected
- Possible Causes:
- Blocked ventilation or cooling fans.
- Extended operation without adequate cooling periods.
- Ambient temperature exceeding recommended operating conditions.
- Solutions:
- Inspect and clean ventilation openings to ensure proper airflow.
- Operate the device within the recommended temperature range.
- Allow the device to rest and cool down between extended uses.
- If overheating persists, seek technical support.
🛠️ General Troubleshooting Tips
- Regular Maintenance: Perform routine inspections and maintenance as outlined in the user manual to ensure optimal performance.
- Proper Storage: Store the device in a clean, dry environment to prevent dust accumulation and moisture exposure.
- Training: Ensure all users are adequately trained in the operation and maintenance of the L10 Light Source.
- Firmware Updates: Keep the device’s firmware up to date to benefit from the latest features and fixes.